Thermometers are a very crucial instrument in the pharmaceutical lab. The output results obtained from the thermometers can impact highly on the whole analysis process done in the lab. To meet the best result a proper SOP for the calibration of Lab thermometers is introduced by the Quality assurance department. In this article, I tried to cover the all parameters of the calibration of lab thermometers.
SOP for calibration of Lab Thermometers:
1. Objective: To lay down a procedure for the Calibration of Lab Thermometers
2. Scope: The procedure of this SOP shall be applicable to the Q.C. Department of ABC Pharmaceuticals
3. Responsibility: Quality Control Chemist
4. Accountability: Quality Control Manager
ALSO READ: Working principle of HVAC system
5. Abbreviations :
S.O.P.: Standard Operating Procedure
QC: Quality Control
QA: Quality Assurance
6. Procedure of calibration of Lab Thermometers:
- Allow the test thermometers to stand at room temperature for 72 hours.
- Place an appropriate amount of oil in a beaker.
- Place the thermometer to be tested in the oil to the depth of the indicated immersion line.
- Place the standard thermometer (Calibrated Thermometer) in the oil to the same depth as the thermometer under test.
- Place the beaker containing oil in a hot place (magnetic), and heat slowly to the desired temperature.
- Read simultaneously the temperature of both thermometers.
- Record readings of both thermometers at various temperatures, for each thermometer to include the temperature at which the thermometer is most frequently utilized.
- Readings should be repeated at least twice and recorded.
Time Interval for Calibration:
Thermometers shall be calibrated at least twice a year and all new thermometers shall be calibrated before use.